Atomic absorption spectrometry for instrumental analysis, following Skoog, Holler and Crouch, Principles of Instrumental Analysis, 7th edition, chapter 9 (with chapter 8 for line widths and temperature). Line wavelengths, ionization energies and masses are from NIST. Flame temperatures are Skoog’s Table 9-1. Where a value is illustrative or a model is schematic, the tab says so. Companion of Atomic Line Spectra, which explains why atoms absorb only narrow lines. The timeline appears only where playing does something; Key moments pause on the events that matter.
The instrument: each part of the light path. Hollow-cathode lamp: ions hit the cathode, atoms are sputtered, they emit the analyte’s line. Solution to atoms: nebulization, the big droplets lost to the drain, desolvation, volatilization, atomization, and the losses to oxides and ions. Graphite furnace: drying, ashing, the atomization peak, clean-out.
Lamp, chopper, flame, monochromator and photomultiplier, after Skoog Figure 9-13. The chopper makes the lamp signal flicker, so the electronics can separate it from the steady emission of the flame; the plot shows this (signal levels illustrative). Atoms spend about 10⁻⁴ s in the light path of a flame.
The lamp is drawn as a real one: octal base with pins, glass envelope with a flat quartz window, getter film, mica disk, lead wires, a hollow cathode cup made of the analyte, and a ring anode. About 300 V ionizes the fill gas (Ne or Ar, 1 to 5 torr); currents of 5 to 15 mA are typical. The visible orange glow is mostly the neon; many useful analyte lines are in the ultraviolet (drawn violet). Pick the cathode metal and the element in the flame: a lamp of one element emits only that element’s lines (NIST: Au 242.795 and 267.595 nm, Fe 248.327 nm and others, Na 588.995/589.592 nm, Ca 422.673 nm), and only matching ground-state lines of the same element can absorb them, so a mismatched lamp reads zero. Lines from excited lower levels (such as Fe 358.1 nm or Au 312.3 nm) are emitted by the lamp but hardly absorbed by flame atoms. Higher current gives more light but a hotter lamp (broader Doppler line) and self-absorption by cool atoms in front of the cathode (self-reversal, Skoog Fig. 9-16); the lamp temperature and self-absorption strength as functions of current are schematic.
Nebulization, desolvation, volatilization and atomization (Skoog Fig. 9-1), with oxides and ions as losses. The flame is drawn as a real air–acetylene flame on a slot burner: a nearly transparent pale-blue sheet with a bright blue ribbon along the slot, tinted by the element sprayed in. The laminar-flow burner sends only the finest mist to the flame; large droplets drain away. Flames: acetylene–air 2100–2400 °C, acetylene–N₂O 2600–2800 °C (Skoog Table 9-1). The absorbance-against-height curves are schematic shapes after Skoog Figure 9-4: an element without a stable oxide gains with height, one forming a stable oxide falls, and one in between peaks in the middle.
The tube is shown cut open so the inside shows: the wall, the sample hole, and the sample sitting on a L’vov platform. 0.5 to 10 µL of sample is dried, ashed and atomized at 2000 to 3000 °C in a graphite tube purged with argon. The programme shown is an example; real programmes depend on the element and matrix. The absorbance pulse uses first-order release (0.3 s) and loss (0.5 s) of atoms, an illustrative model. Because the whole sample is atomized at once in a small volume, detection limits can be about 1000 times lower than with a flame, at the cost of precision and throughput.
Ionization is computed with the Saha equation using ground-state statistical weights (Na and K: g = 2 for the atom, 1 for the ion; alkaline earths: 1 and 2), with the analyte at 10⁻⁶ atm and the charge balance solved including any added K. Check: Willis (Appl. Opt. 7, 1295, 1968) gives 82% (Na) and 98% (K) ionized in an N₂O–acetylene flame at 2950 °C; the module reproduces 81% and 98.5% with his convention (no ground-state weight) and gives 71% and 97% with the textbook weights it uses. Chemical interference of phosphate on Ca and the effects of La, EDTA and the hotter flame are a schematic model. Spectral: V 308.211 nm overlaps Al 308.215 nm (Skoog section 9C-1); Al 309.27 nm is free of it.
Background (scatter and molecular absorption) is broad. The deuterium lamp’s continuum fills the 0.2 nm bandpass, so the narrow atomic line hardly reduces it: it measures background alone. With a magnetic field, the atomic line splits; the module shows the normal Zeeman shift Δλ = λ²eB/(4πmₑc), 16 pm at 1 T for 589 nm, three times the flame line width, with the unshifted component removed by a polarizer. Field off minus field on leaves the atomic absorbance.
Four question types: choosing a flame, fixing ionization, background-correction arithmetic, and matching interferences to fixes. Points: 2 on the first try, 1 after a miss or hint, 0 after giving up.
Lines (NIST, air): Na 588.995, K 766.490, Ca 422.673, Mg 285.213, Sr 460.733, Ba 553.548 nm. Ionization energies (NIST): K 4.3407, Na 5.1391, Ba 5.2117, Sr 5.6949, Ca 6.1132, Mg 7.6462 eV. Doppler widths at the flame or lamp temperature. Constants: CODATA 2022.