Before a mass spectrometer can measure anything, it has to turn neutral molecules into ions, and how that happens decides what the spectrum looks like. In electron ionization, electrons from a hot filament cross the source toward a trap and meet analyte molecules arriving from the GC. A 70 eV electron removes an electron from the analyte (M + e⁻ → M⁺• + 2e⁻) and leaves several eV of excess energy, so M⁺• often breaks into fragments (violet in the 3D view; pale dots are the neutral pieces, which the analyzer never sees). A small positive voltage on the repeller pushes every ion out through the lenses toward the mass analyzer, and the spectrum builds as they arrive.
Below the ionization energy, nothing ionizes. Above it, the yield climbs and levels off well before 70 eV; 70 eV is used because the yield has leveled off there, which makes spectra reproducible. Lower energies give fewer ions and less fragmentation.
The mean free path λ = kBT/(√2·π·d²·P) is the average distance a molecule travels between collisions. At EI pressures (about 10⁻⁶ torr) λ is tens of meters, so in a 1 cm source molecules essentially never collide. Raise the pressure and electrons lose their energy before crossing the source, ions are knocked off course, and at about 1 torr every ion collides hundreds of times, which is the regime of chemical ionization.
EI is a hard method: the fragment pattern is a fingerprint used for library searches, while the molecular ion may be weak or missing. The table lists the main peaks and the mass each one lost from M⁺•. Compare toluene (tropylium, m/z 91, from loss of H), acetone (CH₃CO⁺, m/z 43, loss of CH₃) and pyridine (loss of HCN).
Identify an analyte from its spectrum, name a neutral loss, find the ionization threshold, and compute a mean free path.
Source about 1 cm across (drawn 8 units wide), 200 °C, collision diameter 0.37 nm; electron mean free path taken as 4√2 times that of the molecules. EI spectra, ionization energies and the efficiency curve are those of the Ionization Explorer, carried over unchanged. Motion in the 3D view is slowed enormously and particles are drawn much larger than molecules. Lecture-specific content will be added when the lecture notes arrive.