Electrospray ionization (ESI)

capillary: +3 to 5 kV inlet: 0 V (ground) +−high-voltage supply electric field (+ to ground) to MS warm N₂ drying gas Taylor cone jet charged droplets shrink as solvent evaporates gas-phase ions

The whole process. A high-voltage supply holds the metal capillary at +3 to 5 kV and the inlet at ground, so the electric field points from the tip toward the inlet and charges the liquid as it leaves the tip. The liquid forms a Taylor cone and a jet that breaks into a spray of charged droplets carrying analyte molecules. Warm drying gas evaporates the solvent as the droplets drift toward the inlet; they shrink and split until the analyte leaves as gas-phase ions that enter the mass spectrometer.

metal capillary: + electrode, +3 to 5 kV + + + + + + + + + + + + + + + + + + inlet: 0 V +−high-voltage supply electric field (+ to ground) Taylor cone jet breaks into charged droplets positive ions move with the field to the tipnegative ions move back toward the positive capillary wallanalyte (phenylalanine)

Taylor cone. The metal capillary is the positive electrode and the inlet is at ground, so the field points from the tip toward the inlet. Positive ions in the liquid move with the field to the liquid surface at the tip, while negative ions move the other way, back toward the positive capillary wall. The excess positive charge at the surface pulls the liquid into a cone, and a thin jet from its tip breaks into droplets that carry that charge along with dissolved analyte.

evaporation fission field-assistedion desorption [M + 3H]3+ Dashed circles mark each stage; after the textbook scheme, not to scale.

Evaporation and fission. Solvent evaporates but the charge stays, so the droplet shrinks until it reaches the Rayleigh limit, where charge repulsion matches surface tension. It then undergoes fission and sheds small, highly charged droplets. The remaining droplet keeps shrinking until the electric field at its surface is strong enough to lift the analyte out as an ion (field-assisted ion desorption).

Small ions: ion evaporation Larger molecules: charged residue the field at the surface pushes an [M + H]+ ion out [M + 3H]3+ the droplet dries out; its protons stay on the basic N atoms

Ion release. Two models explain the last step. Small ions are pushed out of a tiny, highly charged droplet by the field at its surface (ion evaporation, Iribarne and Thomson). Larger molecules such as peptides and proteins stay until the last solvent is gone, and the remaining protons stay on their basic sites (charged residue, Dole).

droplet with 2 charges [M + 2H]2+ droplet with 3 charges [M + 3H]3+ droplet with 4 charges [M + 4H]4+ m/z one molecule, three charge states, three peaks 4+ 3+ 2+

Why so many charge states? When a droplet dries out, its molecule keeps the charges the droplet still holds, up to the number of basic sites that can take a proton; this peptide has four (the N-terminal amine and three side-chain amines). Droplets differ in size and charge, so identical molecules end up with different numbers of protons: [M + 2H]2+, [M + 3H]3+ and [M + 4H]4+. Each charge state gives its own peak. A protein with dozens of basic sites spreads over many more, such as +11 to +17 for cytochrome c.

inlet +12 +15 +18 one protein, differentnumbers of protons 700 800 900 1000 1100 m/z +11 +12 +13 +14 +15 +16 +17 cytochrome c, M = 12,360 Da

Into the analyzer. The same protein arrives with different numbers of protons, so one molecule gives a series of peaks. Each peak appears at m/z = (M + n·mH)/n, so more protons means lower m/z. Two neighboring peaks are enough to find n and M.

Drag to turn the view. Scroll or pinch to zoom. Double-click to reset.

How ESI makes ions

How electrospray ionization makes ions. Step through how a solution becomes gas-phase ions: from the Taylor cone, through evaporation and Coulomb fission, to multiply charged ions entering the mass spectrometer. Use “Show” to play the whole process or one step at a time; the view moves to each step.

The whole process. A high-voltage supply holds the metal capillary at +3 to 5 kV and the inlet at ground, so the electric field points from the tip toward the inlet and charges the liquid as it leaves the tip. The liquid forms a Taylor cone and a jet that breaks into a spray of charged droplets carrying analyte molecules. Warm drying gas evaporates the solvent as the droplets drift toward the inlet; they shrink and split until the analyte leaves as gas-phase ions that enter the mass spectrometer.

Taylor cone. The metal capillary is the positive electrode and the inlet is at ground, so the field points from the tip toward the inlet. Positive ions in the liquid move with the field to the liquid surface at the tip, while negative ions move the other way, back toward the positive capillary wall. The excess positive charge at the surface pulls the liquid into a cone, and a thin jet from its tip breaks into droplets that carry that charge along with dissolved analyte.

Evaporation and fission. Solvent evaporates but the charge stays, so the droplet shrinks until it reaches the Rayleigh limit, where charge repulsion matches surface tension. It then undergoes fission and sheds small, highly charged droplets. The remaining droplet keeps shrinking until the electric field at its surface is strong enough to lift the analyte out as an ion (field-assisted ion desorption).

Ion release. Two models explain the last step. Small ions are pushed out of a tiny, highly charged droplet by the field at its surface (ion evaporation, Iribarne and Thomson). Larger molecules such as peptides and proteins stay until the last solvent is gone, and the remaining protons stay on their basic sites (charged residue, Dole).

Why so many charge states? When a droplet dries out, its molecule keeps the charges the droplet still holds, up to the number of basic sites that can take a proton; this peptide has four (the N-terminal amine and three side-chain amines). Droplets differ in size and charge, so identical molecules end up with different numbers of protons: [M + 2H]²⁺, [M + 3H]³⁺ and [M + 4H]⁴⁺. Each charge state gives its own peak. A protein with dozens of basic sites spreads over many more, such as +11 to +17 for cytochrome c.

Into the analyzer. The same protein arrives with different numbers of protons, so one molecule gives a series of peaks. Each peak appears at m/z = (M + n·mH)/n, so more protons means lower m/z. Two neighboring peaks are enough to find n and M.

Schematic, not to scale. Atoms: C dark gray, H white, N blue, O red; red pins are protons. Mechanism after Fenn et al., Science 246, 64 (1989).

Spray and droplets

A Taylor cone forms only above an onset voltage that depends on the surface tension of the liquid (lower with more organic solvent), the capillary radius and the distance to the inlet; well above onset the spray becomes unstable and may discharge. The plot follows one droplet: solvent evaporates, the charge stays, and each time the droplet reaches the Rayleigh limit, qR = 8π√(ε₀γr³), it splits. Warmer drying gas and more organic solvent speed evaporation (the evaporation constant here is illustrative). The charging setting compares the denatured charge states of Fenn et al. (1989) with the fewer charges a compact, native protein keeps.

Charge states: n and M

ESI spectra: charge, mass and prediction. Work out the charge and mass from two neighboring peaks, or pick a protein and see where its peaks should appear.

Positive-ion mode with protons as the charges; mH = 1.007276 Da. Peak positions are calculated; peak heights are illustrative.

Masses and charge ranges from Fenn et al., Science 246, 64 (1989): Fig. 9 for the worked examples, Table 1 for the predictions.

ESI versus MALDI

The same protein by both methods: electrospray spreads it over many charge states at lower m/z, while MALDI puts nearly everything into one peak at M + 1.

ESI challenge

Find the charge from two neighboring peaks, find the mass, predict where a charge state appears, and decide which peak carries more protons.

Settings used

Spray model, solvents and protein charge ranges are those of the ESI & MALDI Explorer, carried over unchanged; the step texts are the original ESI animation’s, and the worked examples follow the original charge calculator. The 3D view is schematic, not to scale. Lecture-specific content will be added when the lecture notes arrive.