The whole process. A high-voltage supply holds the metal capillary at +3 to 5 kV and the inlet at ground, so the electric field points from the tip toward the inlet and charges the liquid as it leaves the tip. The liquid forms a Taylor cone and a jet that breaks into a spray of charged droplets carrying analyte molecules. Warm drying gas evaporates the solvent as the droplets drift toward the inlet; they shrink and split until the analyte leaves as gas-phase ions that enter the mass spectrometer.
Taylor cone. The metal capillary is the positive electrode and the inlet is at ground, so the field points from the tip toward the inlet. Positive ions in the liquid move with the field to the liquid surface at the tip, while negative ions move the other way, back toward the positive capillary wall. The excess positive charge at the surface pulls the liquid into a cone, and a thin jet from its tip breaks into droplets that carry that charge along with dissolved analyte.
Evaporation and fission. Solvent evaporates but the charge stays, so the droplet shrinks until it reaches the Rayleigh limit, where charge repulsion matches surface tension. It then undergoes fission and sheds small, highly charged droplets. The remaining droplet keeps shrinking until the electric field at its surface is strong enough to lift the analyte out as an ion (field-assisted ion desorption).
Ion release. Two models explain the last step. Small ions are pushed out of a tiny, highly charged droplet by the field at its surface (ion evaporation, Iribarne and Thomson). Larger molecules such as peptides and proteins stay until the last solvent is gone, and the remaining protons stay on their basic sites (charged residue, Dole).
Why so many charge states? When a droplet dries out, its molecule keeps the charges the droplet still holds, up to the number of basic sites that can take a proton; this peptide has four (the N-terminal amine and three side-chain amines). Droplets differ in size and charge, so identical molecules end up with different numbers of protons: [M + 2H]2+, [M + 3H]3+ and [M + 4H]4+. Each charge state gives its own peak. A protein with dozens of basic sites spreads over many more, such as +11 to +17 for cytochrome c.
Into the analyzer. The same protein arrives with different numbers of protons, so one molecule gives a series of peaks. Each peak appears at m/z = (M + n·mH)/n, so more protons means lower m/z. Two neighboring peaks are enough to find n and M.