The inductively coupled plasma for instrumental analysis: plasma as a state of matter, how a radio-frequency coil heats argon to about 10,000 K, what happens to a sample inside, why the ICP beats a flame, the échelle spectrometer, and ICP-MS. Following Skoog, Holler and Crouch, Principles of Instrumental Analysis, chapters 10 and 11, and R. Thomas, “A beginner’s guide to ICP-MS”, Spectroscopy (2001). Ionization uses the Saha equation; excitation the Boltzmann equation; lines and ionization energies are from NIST. Values marked schematic are drawn to show the idea, not measured.
Plasma: argon melts, boils, the first ions at 6,000 K, an ICP’s hottest zone at 10,000 K, lightning at 28,000 K. The torch: coil on with nothing happening, the spark, the field reversing, the collision avalanche, the self-sustaining plasma, the nearly motionless ions. Inside the plasma: a droplet’s journey. ICP-MS: ions through the cones and the quadrupole.
Argon at 1 atm. Melting 83.8 K (the triple point 83.8058 K is a fixed point of the ITS-90 temperature scale) and boiling 87.30 K. Ionization from the Saha equation (ionization energy 15.7596 eV, NIST; Ar⁺ ground term 2P with levels at 0 and 0.177 eV): about 2.4 × 10⁻⁵ ionized at 6,000 K, 1.6 × 10⁻³ at 8,000 K, 2% at 10,000 K, and almost fully (singly) ionized at 28,000 K, the lightning temperature given by NOAA’s National Weather Service (about 50,000 °F). The 3D box shows 216 atoms with the ionized number to scale.
Three concentric quartz tubes; a water-cooled copper load coil carries radio-frequency current at 27.12 or 40.68 MHz with 0.5 to 2 kW. Its alternating magnetic field induces a circular electric field in the gas (Faraday’s law). Cold argon has no free charges, so nothing happens until a Tesla-coil spark seeds electrons; the field then drives them back and forth around the torch, reversing every half-cycle, and their collisions with argon atoms heat the gas and ionize more atoms (an avalanche) until the plasma sustains itself. An argon ion is 72,821 times heavier than an electron, so the ions hardly move. The animation is slowed down billions of times; the electron count and the gas-temperature curve are schematic.
Zone temperatures after Thomas (2001, Fig. 4): about 10,000 K in the induction zone, falling through 8,000, 7,500, 6,500 and 6,000 K up the plume. Plasma gas 12 to 17 L/min, auxiliary about 1 L/min, carrier (nebulizer) gas about 1 L/min. The sample is desolvated, vaporized, atomized, excited and ionized, spending about 2 ms at 6,000 to 8,000 K (Skoog). Radial viewing looks across the plasma; axial viewing looks down its length.
Excited fraction N*/N₀ = 3 exp(−hc/λkT) for each element’s resonance line (NIST). Ionization in the flame from the Saha equation with the analyte at 10⁻⁶ atm (as in the AAS module); in the ICP with the electron density from ionized argon at the chosen temperature. Advantages listed after Skoog chapter 10.
An échelle grating works in high orders; a prism separates them, folding the spectrum onto a CCD. Line wavelengths are from NIST; the order constant (m·λ ≈ 22,500 nm) is schematic.
Ions formed in the plasma pass the sampler and skimmer cones into vacuum and a quadrupole mass filter (see the Quadrupole Mass Filter module). Ionization at 7,500 K from the Saha equation. Classic interferences: ⁴⁰Ar⁺ on ⁴⁰Ca⁺ and ⁴⁰Ar¹⁶O⁺ on ⁵⁶Fe⁺. Bar heights are schematic.
Five question types on states of matter, RF heating, the torch, ICP versus flame, and ICP-MS. Points: 2 on the first try, 1 after a miss or hint, 0 after giving up.