Inside the Column

Drag to turn the view. Scroll or pinch to zoom. Double-click to reset.

Ride with a molecule

Put yourself inside the column. In HPLC and packed GC you weave between the packing particles, and no two routes are the same length; in a capillary column you bounce off the wall and dip in and out of the stationary film while the flow carries you toward the detector. Use the view buttons to sit in the molecule’s eyes, follow just behind it, or pull back and watch the whole band. The band behind you widens as σ = √(H·x), and the plate height H is exactly what the Van Deemter curve on the right is telling you.

Interactive Chemistry · Ali Shafiee · University of Florida

Every band of molecules spreads out as it moves through a column, and the more it spreads, the fewer compounds the column can separate. The plate height H measures that spreading: smaller H means sharper peaks. The Van Deemter equation splits H into three causes:

H = A + B/u + (CS + CM)·u

Pick a column type at the top. Change parameters in the panel (Flow, A, B, C). Under each parameter is the equation it enters, with the parameter in red and today’s numbers plugged in, so you can see whether it sits in a numerator or a denominator, is squared, or is raised to a power. Tap any dashed value to type an exact number, which is handy for practice problems. H and the bar of contributions update instantly. Switch views to see the graph, a 3D column, close-ups of each term, or every number and coefficient. On the graph, drag sideways to change the velocity.

Molecule size and carrier gas

In the Molecules view, the analyte is drawn in its real shape (ball-and-stick, with a faint shell showing its size as other molecules feel it) among carrier gas molecules, which bounce off it. The faint dots are copies of the analyte that all started at the same point, and the ring shows how far they have spread: faster diffusion means a ring that grows faster. The animation is in slow motion and the gas is drawn much denser than it really is so the collisions are visible.

In the B tab you can pick a real analyte and turn on Compare to set it against a second analyte or a second carrier gas (or solvent). The Molecules view shows each analyte in its real shape being knocked around by carrier gas molecules. Bigger analytes and bulkier, heavier carrier gases mean slower diffusion (smaller DM).

The carrier gas does not push big molecules more slowly: at the same linear velocity u, every analyte molecule in the gas moves down the column at the same speed. What the gas changes is how fast the analyte diffuses, which changes B and CM, the shape of the curve and the optimum velocity. In a real GC column, bigger molecules also have higher boiling points and larger k, so they elute later too; here both use the same k so you can see the diffusion effect on its own.

A · Multiple paths

A · Multiple paths (eddy diffusion)

In a packed column, molecules weave around particles. Some find nearly straight routes, others take long detours, so they arrive at different times. Bigger particles (dp) and uneven packing (λ) make paths more varied. Flow speed doesn’t change this, so A is a flat line. Open tubular (capillary) columns have no particles: A = 0.

B · Diffusion

B · Longitudinal diffusion

Molecules wander forward and backward along the column the whole time they are in it. Slow flow means more time to wander, so this term is B/u. Faster diffusion (larger DM: light carrier gas, small analyte, high temperature, low pressure) makes B larger. Diffusion in liquids is about 10,000 times slower than in gases, so B matters little in HPLC except at very low flow.

CS · Mass transfer in the stationary phase

Molecules that dissolve into the stationary phase stop moving until they diffuse back out, while the others are carried ahead. A thicker film (df) or slower diffusion in it (DS) keeps them longer, and faster flow carries the others farther meanwhile, so the term is CS·u. In HPLC the bonded phase is only nanometers thick, so CS is usually tiny.

CM · Mass transfer in the mobile phase

Flow is fastest in the middle of a channel and nearly zero at the walls. Molecules leave a fast or slow stream only by diffusing sideways. Wider channels (bigger particles, or a wider capillary) and slower diffusion stretch the band more, and faster flow makes it worse: CM·u. This is why small particles and narrow capillaries give sharp peaks.

A = 2λdp (packed), A = 0 (open tube). B = 2γDM, with γ = 1 in an open tube. CS = (2/3)·k/(1+k)²·df²/DS. CM = f(k)·r²/DM for an open tube of radius r (Golay), with f(k) = (1+6k+11k²)/(24(1+k)²); for packed beds CM = ω·dp²/DM with ω = f(k)/4.

DM in gases follows the Fuller equation (rises as T1.75, falls with pressure, larger for light carrier gases and small analytes). DM in liquids follows the Wilke–Chang equation (rises with temperature, falls with solvent viscosity, smaller for large analytes).

uopt = √(B/C) and Hmin = A + 2√(BC). A column of length L has N = L/H plates, tR = (L/u)(1 + k), and a peak leaves with a width of 4σ = 4tR/√N.

Values are typical textbook magnitudes chosen to show realistic trends. In a real column, temperature also changes k; here k has its own slider so you can see each effect separately. The 3D view and close-ups are magnified and not to scale; the film thickness is exaggerated so you can see it.

C · Mass transfer

Every time the molecule enters the stationary phase it stops while the flow carries the rest of the band on, and it only catches up when it comes back out. The faster the flow, the further behind it falls, so this term grows with velocity: C·u. A thinner film, a smaller particle or faster diffusion in the stationary phase all reduce it.

The curve

A is flat, B/u falls as the flow speeds up and C·u rises, so their sum has a minimum. That minimum is the optimum velocity, uopt = √(B/C), where the plate height is Hmin = A + 2√(BC). Running slower than the optimum costs you diffusion; running faster costs you mass transfer. The breakdown under the curve shows how much each term contributes at your current settings.

Challenge

Four kinds of rounds: find the optimum velocity, say which term dominates, explain why running too slowly broadens peaks, and compare packed and capillary columns.

Settings used

Column types, parameters, analytes, carrier gases and the plate-height model are those of the Van Deemter Explorer, carried over unchanged. The 3D column is schematic: particle sizes, molecule sizes and speeds are not to scale, and the motion is slowed so the three effects are visible. Lecture-specific content will be added when the chromatography lecture notes arrive.