Step through how a laser pulse turns matrix crystals into gas-phase ions: sample spot, laser pulse, ablation, proton transfer, and extraction into a time-of-flight analyzer. Use “Show” to play the whole process or one step at a time; drag the timeline to move through it.
The whole process. A short UV laser pulse hits a spot of matrix crystals that contain the analyte. The matrix absorbs the light and erupts into a plume, where protons pass from matrix to analyte. The target plate is held at +20 kV and the grid at ground, so positive ions are pushed into the time-of-flight tube, where lighter ions reach the detector first.
Sample spot. The analyte is mixed with a large excess of a UV-absorbing matrix, here 2,5-dihydroxybenzoic acid (DHB), typically about 1,000 to 10,000 matrix molecules per analyte molecule. As the droplet dries on the metal target, the matrix crystallizes and traps the analyte molecules, keeping them apart from one another.
Laser pulse. A pulse of UV light lasting a few nanoseconds is absorbed by the matrix, not the analyte. Matrix molecules near the surface are excited (M + hν → M*), and the absorbed energy heats the top of the crystal almost instantly.
Ablation. The sudden heating blows the top layers off the crystal as a dense, expanding plume of matrix molecules and small clusters that still hold analyte molecules. Because the matrix takes up most of the energy, the analyte goes into the gas phase without breaking apart.
Proton transfer. In the hot, dense plume, an excited matrix molecule gives a proton to another matrix molecule: M* + M → MH⁺ + (M−H)⁻. The protonated matrix then hands its proton to the analyte, which holds a proton more strongly: MH⁺ + A → AH⁺ + M. Here M is the matrix and A the analyte.
Extraction and time of flight. Clusters lose their remaining matrix. The target at +20 kV pushes positive ions toward the grounded grid, while negative ions such as (M−H)⁻ are pulled back to the plate. In the flight tube, lighter ions arrive first. Unlike ESI, MALDI gives mostly singly charged ions, so the main peak is AH⁺, with only a small doubly charged peak.
Schematic, not to scale. Atoms: C dark gray, H white, N blue, O red; red pins are protons, blue pins mark negative charge. Matrix shown: 2,5-dihydroxybenzoic acid. Reactions as on the course slide (M = matrix, A = analyte).
The matrix absorbs the UV photon (E = hc/λ: 3.68 eV at 337 nm from a nitrogen laser, 3.49 eV at 355 nm from a tripled Nd:YAG laser) through its aromatic ring; the analyte does not. Below a threshold fluence nothing desorbs; above it the ion signal rises very steeply, and too much laser fragments the analyte and broadens the peaks. The matrix suits some analytes better than others: CHCA for peptides, sinapinic acid for large proteins, DHB for both. The analyte must be diluted in the matrix, about 1,000 to 10,000 matrix molecules per analyte molecule: too little matrix and the analyte is not isolated, too much dilution and few analyte molecules are in the irradiated spot.
MALDI gives mostly singly charged ions, so the main peak is [M+H]⁺, with a small [M+2H]²⁺ and, for proteins, a [2M+H]⁺ dimer; peptides also show Na⁺ and K⁺ adducts. Matrix ions crowd the low-m/z end. Zoom on [M+H]⁺ to see the peak width set by the time-of-flight resolution, which falls for heavier ions.
The same protein by both methods: MALDI puts nearly everything into one peak at M + 1, while electrospray spreads it over many charge states at lower m/z.
Find the main peak, choose a matrix, locate the doubly charged peak, and compare flight times.
Matrices, analytes, yields, resolution and the ESI charge states are those of the ESI & MALDI Explorer, carried over unchanged; the step texts are the original MALDI animation’s. Flight times for a 1.0 m tube at 20 kV. Lecture-specific content will be added when the lecture notes arrive.