Quadrupole mass filter

Drag to turn the rods. Scroll or pinch to zoom. Double-click to reset.

The rods and the field

Four parallel rods; opposite rods are connected. The x rods get φ0 = U + V cos ωt and the y rods −φ0 (slide 9). Inside, the potential is φ = φ0(x² − y²)/r0², so the field grows linearly with distance from the axis (slide 6) and the equipotentials are saddles (slide 7). The RF runs slowly in this part, about 5 seconds per cycle, so the rods and the potential slice change color gradually as the polarity reverses. The two plots show the potential along x and along y: one direction confines a positive ion while the other pushes it out, and half a cycle later they swap. Send an ion through with DC only, RF only, or both to see slide 10: DC alone drags the ion into the negative y rods, RF alone makes it oscillate. The end view can show round rods (real) or hyperbolic rods (ideal, slide 8).

One ion’s journey

Four rods carry a DC voltage U plus an RF voltage V cos ωt. Opposite rods are connected, so the potential between the two pairs is Φ₀ = U + V cos ωt. Only ions of the right m/z have a stable path and reach the detector.

Choose an ion, set U and V, and press play. The path is calculated from the equations of motion, not drawn by hand. Too light, and the AC field throws the ion into the x rods; too heavy, and the DC field drags it into the y rods.

The dot on the stability diagram is the ion’s working point (q, a). Inside the shaded region the motion is stable in both x and y. Drag the dot to change U and V directly.

Filtering a mixture

Several kinds of ions enter together. They all feel the same U and V, but each has its own a and q, so each sits at its own point on the same line through the origin of the diagram (slides 20, 23–24). Move U and V, or tap “Keep only” for one ion: lighter ions are thrown out by the AC field and hit x rods; heavier ones are dragged off by the DC field and hit y rods. With U = 0 every ion with q below 0.908 passes (RF-only mode).

Scanning

U and V rise together, and the working point climbs the scan line through each ion’s stable region in turn, giving one peak per ion (slides 18–21). With a fixed U/V the scan line is straight and passes through the origin: a steeper line cuts a thinner slice of each region, so peaks are narrower and resolution higher, but fewer ions pass (slides 22, 25). At the tip, a/q = 0.336 and U/V = 0.168 (slide 26). With a fixed U/V each peak’s width grows in proportion to m, so R = m/Δm stays the same at every mass. Instruments normally raise U/V slightly as the mass increases so that every peak is about 1 m/z wide (unit resolution); then R grows in proportion to m, as on slide 4. Both modes are available. With U = 0 (RF only) there are no peaks at all (slide 27). Peak shapes come from flying a batch of ions with different starting positions and RF phases at every step of the scan.

Triple quad

Q1 is a mass filter set to pass only the precursor (here beauverolide, [M+H]⁺ m/z 488). q2 is RF only, an ion guide filled with collision gas: ions collide with the gas and fragment (CID). Q3 scans like a single quadrupole, so each fragment reaches the detector in turn: a product ion scan (slides 28–30). Switch the gas off to see q2 simply guide the precursor through. The fragments and their intensities come from the MS² example on slide 52, which was recorded on an ion trap; a triple quad gives a similar set of fragments, with intensities that depend on the collision energy. The triple quad’s peaks use the ideal stability windows.

Mass filter challenge

A game built on the same physics. Each round draws a few real ions from a hidden bank. In “Guess the winner” the voltages are fixed: work out a and q for every ion from the Conditions box (U, V, r0, the RF frequency and the equations are all given), find the one inside the stable region, and tap it before firing; the diagram shows the ions’ real points only after they fly. In the other rounds you set U and V so that only the target reaches the detector: separating neighbors one mass unit apart needs a scan line close to the tip. Some rounds include two ions with the same nominal m/z; then the right answer is “These can’t be separated”, and the game shows the resolving power that would be needed.

Settings used

Rod spacing r₀ = 4.0 mm, RF frequency 1.00 MHz, rod length 15 cm, ion energy 5 eV, singly charged ions starting 0.1 r₀ off the axis. In the 3D view the rods are shortened to fit the screen.

Equations follow the lecture: a = 4eU/(m r₀²ω²), q = 2eV/(m r₀²ω²), where Φ₀ = U + V cos ωt is the potential between the rod pairs. Stability limits are the Mathieu characteristic curves (apex a = 0.237, q = 0.706; edge q = 0.908 at a = 0).