Two curved plates, the outer at +V and the inner at −V, make a radial field that bends ions around the sector (slides 18–20). From qE = mv²/r and ½mv² = qVs, r = 2Vs/E: mass and charge cancel, so the sector sorts ions by kinetic energy per charge, not by mass. With “One ion, three energies”, ions 2% slower and 2% faster than the rest land about 6 mm on either side of the centre of the focal plane, and ions of the same energy entering at different angles meet again there (slide 20). Read every energy at once with the array detector, or move a slit to pick one energy; sweeping the plate voltage brings each energy to the slit in turn, which is the kinetic-energy spectrum of slide 19. With “Three masses, one energy”, m/z 28, 720 and 5734 land on exactly the same spot. In multi-collector ICP-MS the electric sector is used this way, as an energy filter in front of the magnet.
In a magnetic field, qvB = mv²/r gives r = mv/qB (slides 7–8). Every ion leaves the source with the same energy, so momentum and radius grow as √m: in a field set for Ar⁺ (25.0 cm), N₂⁺• follows 20.9 cm, O₂⁺• 22.4 cm and CO₂⁺• 26.2 cm. These modest differences in curvature, set by m/z = B²r²e/2Vs (slide 9), bring each ion to its own focus on the focal plane, which is tilted to the beam. Collect them as on slides 9, 10 and 16: a row of cups reads every ion at once, a movable detector reads one position at a time, or a fixed detector reads each ion in turn while B is scanned. The signal is drawn on a log scale because air is 78% N₂ and only 0.04% CO₂. The angle spread refocuses at the focal plane; an energy spread blurs it (slides 13–15).
The beam holds three krypton isotopes, ⁸²Kr⁺, ⁸³Kr⁺ and ⁸⁴Kr⁺, about 1.2% apart in mass, each with a spread of energies and angles. With the magnet alone, each isotope is smeared by its energy spread into a band about 5 mm wide, wider than the 3 mm between isotopes, so they merge. Put an electric sector in front (slides 21–22): it spreads energies one way and the magnet spreads them the other way by the same amount, so every ion of one isotope lands on one point, whatever its energy or angle. Different masses are not refocused, because the electric sector does not disperse mass at all: the three isotopes stay 2.9 mm apart and come out as three sharp, separate peaks. Nier–Johnson (90° electric then 60° magnetic, slide 24) does this at one exit slit; Mattauch–Herzog (31.8° then 90°, slide 26) does it for every mass along a straight focal plane. Sideways differences are drawn 20× larger in the 3D view; the plot of landing positions is to true scale.
A close-up of the collector block of a multi-collector ICP-MS (slide 2), with numbers from the Thermo Neptune: nine Faraday cups, eight of them movable along the focal plane, and a mass dispersion of about 810 mm, so ⁸⁶Sr, ⁸⁷Sr and ⁸⁸Sr sit 9.3 mm apart and every cup reads at the same moment. The plasma flickers, but all isotopes rise and fall together, so the ratio stays steady; one collector stepping between isotopes scatters about a hundred times more. Iron is harder: ⁵⁶Fe⁺ and ⁴⁰Ar¹⁶O⁺ are 0.0224 u apart, 0.32 mm on the focal plane. At high resolution (narrow entrance slit, R ≈ 8500 on the peak edge) the ⁵⁶Fe peak has a flat shoulder free of ArO, where labs make the measurement (Weyer and Schwieters, 2003); at low resolution (R ≈ 500) the two always overlap. ⁸⁷Sr and ⁸⁷Rb differ by 0.0003 u and would need R ≈ 290 000, beyond any sector, so rubidium is removed chemically first.
Three kinds of rounds, all solvable by hand from the Conditions box: the field B that brings a given ion to the detector (m/z = B²r²e/2Vs), the voltage an electric sector passes or which ions it passes (r = 2Vs/E), and the resolving power R = m/Δm needed to separate real interferences such as CO⁺ and N₂⁺ or ⁴⁰Ca⁺ and ⁴⁰Ar⁺.
Ion paths are traced step by step through the real fields: a radial electric field falling as 1/r between the plates, and a uniform magnetic field between the pole pieces. Ions entering the electric sector off the central radius gain or lose energy in its fringe field. Peak widths and resolving power come from first-order ion optics, checked against the traced paths. Electric sectors: radius 0.30 m, plate gap 3.0 cm. Magnets: radius 0.25 m (Mattauch–Herzog 0.20 m). Accelerating voltage 3.0 to 5.0 kV. Exact isotope masses and natural abundances. Idealizations: sharp field edges, no space charge, singly charged ions. Collector cup spacing is drawn 12 times larger than real. Slide notes: on slide 8 the middle step should read m²v² = 2qVsm; on slide 17, both 60° and 180° magnets disperse masses and focus directions (the angle sets where the focus lies); an electric sector selects kinetic energy per charge and a magnet momentum per charge.