HexaMole mascotHexaMoleInteractive Chemistry

Time-of-flight mass analyzer

Drag to turn the view. Scroll or pinch to zoom. Double-click to reset.

Reflectron TOF

The design used in labs today (slides 24–27). Every ion leaves the source with the same kinetic energy, zeVs = ½mv² (slide 7), so lighter ions move faster: three ions with very different masses (m/z 28, 720 and 5734) arrive in a ratio of about 1 : 5 : 14, because flight time grows as √(m/z) (slide 8). They fly down the field-free tube, turn around in the ion mirror and fly back to a detector beside the source. The mirror is a stack of ring electrodes, each a fixed step higher in voltage than the one before, so the potential rises evenly with depth. An ion stops where the potential equals its energy per charge; the panel on the right shows that depth, the same for every ion, whatever its mass. Change the voltage step to move the mirror off focus, or compare with a straight tube of the same length.

Inside the reflectron

Why the stepped ring voltages fix the energy spread (slides 21–25). Three ions of the same m/z leave with different energies (±8% by default, much larger than in a real instrument, so the difference in depth is easy to see). In the tube the faster one pulls ahead. In the mirror it has more energy, so it climbs further up the potential staircase, goes deeper and spends longer turning around. With the right voltage step, the extra time in the mirror exactly cancels its lead, and all three reach the detector together. For a single-stage mirror this happens when the drift length is four times the turning depth (L = 4d). The position-against-time plot shows the lines crossing at the detector; the spectrum compares the peak with and without the mirror.

Linear TOF (the original design)

The first TOF instruments had no mirror: a straight field-free tube with the detector at the end. The flight-time law is the same, tD = D(m/2zeVs)½, and the master equation m/z = 2eVst²/D² turns a measured time into m/z (slides 8–9; slide 11 gives 8.52 µs for CO⁺• and 8.67 µs for HCO⁺ at 2.000 kV over 1.00 m). But nothing makes a faster ion wait, so ions of the same m/z with slightly different energies simply arrive apart, and the peaks are wide (slide 26: R ≈ 600 linear versus 3400 with a reflectron). A metastable ion that breaks apart in the tube keeps its speed, so its fragment arrives with it (slide 20).

Separating neighbors

The time between two ions depends on √m₂ − √m₁, not on m₂ − m₁ (slide 13), so the same mass difference gives a smaller gap at higher mass (slides 14–16). The diagram is slide 16 with the current peak width drawn across it: neighbors stay resolved only while the gap is wider than a peak. The ability to separate neighbors falls with mass, even though the resolving power R = m/Δm does not fall in the same way; what grows is the peak width in m/z.

Why peaks broaden

A laser pulse lasts several nanoseconds, so ions start at different moments; this part shows the first 250 ns near the source, where the staggered starts are visible. That offset survives the whole flight, at any tube length (slides 17–19), and blurs neighbors together. Pulsed extraction switches the field on after the pulse, so every ion starts together and the peaks shrink (slide 23). Ions born at different depths get different energies; the deeper ones overtake the others at the first focal point F1, about twice their distance from the exit (slides 21–22).

TOF challenge

Three kinds of rounds, all solvable by hand from the Conditions box: identify an ion from its flight time (m/z = 2eVst²/D²), compute the time gap between two ions, and choose the shortest tube that separates two neighbors. The ions come from a hidden bank of real ions, and every answer is explained with numbers.

Settings used

Single-field source, gap 5 mm; e = 1.602177×10⁻¹⁹ C, 1 u = 1.660539×10⁻²⁷ kg; ions singly charged; detector response 1 ns. Each species is flown as a batch of 140 ions with the chosen spreads, and peak widths (FWHM) come from those simulated arrival times. Reflectron: single-stage mirror of 20 ring electrodes with equal voltage steps (a homogeneous field), both legs of the flight path the same length; small entry angles are ignored, and the return path is drawn at an exaggerated angle. Idealizations: no initial thermal velocities, ideal grids, pulsed extraction modeled as removing the birth-time spread only. Equations: Wiley and McLaren, Rev. Sci. Instrum. 26, 1150 (1955); single-stage reflectron focusing L = 4d, Mamyrin et al. (1973), Cornish, Johns Hopkins APL Tech. Dig. 20 (1999). Slide notes: on slide 14 the lower annotation (next to 490 → 500) should read “large masses → small difference in √m → ions separate poorly”; on slide 15 the charge should read 1.6022×10⁻¹⁹ C.